Open original FDA record ↗Recall record
- FDA identifier
- Z-0973-2014
- Company in FDA record
- Medtronic Navigation, Inc.
- FDA classification date
- Feb 28, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Review of on-system electronic product labeling and Instructions for comply with applicable Use in response to a customer inquiry uncovered gaps in compliance FDA standards with the requirements for information to be provided to users in 21CFR 1020.30(h)
Source facts
- Action
- Medtronic Navigation Inc. issued Notiifcation on 3/4/14 and included thee errata sheet to all affected customers updating the missing and incorrect information and ensure that labeling is now in compliance with 21 CFR Part 1020.30(h). This will be followed by a software update that contains the same information on the errata sheet, which serves to correct the electronic version of the Instructions for Use. The paper errata sheet will be sent via traceable means and the software upgrade will be performed by Medtronic Service either through remote access or by site visit of Medtronic service personnel. For further questions please call (978) 698-6000.
- Initiated
- 2013-07-25
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2019-04-17
- Distribution
- USA (nationwide) Distribution
- Recall number
- Z-0973-2014
- Recall status
- Terminated
- CDRH record ID
- 125338
- FDA classified
- 2014-02-28
- Recall event number
- 67363