Skip to content

Public FDA record

Z-0973-2019: Stryker Sustainability Solutions (SSS) reprocessed Arthroscopic Shavers Arthroscopic shavers can be use

Recall record Z-0973-2019 for Stryker Sustainability Solutions. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0973-2019
Company in FDA record
Stryker Sustainability Solutions
FDA record creation date
Mar 5, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Affected lots of reprocessed arthroscopic shavers may display the incorrect catalog number and speed settings when connected to the generator, require manual mode for use on CORE generators, or display an error message on Crossfire 1 and 2 generators which does not allow the device to be used.

Source facts

Action
On July 26, 2018, Stryker Sustainability Solutions sent an "URGENT MEDICAL DEVICE RECALL" Customer Notification Letter to affected consignees asking them to follow the instructions in the letter. The letter also included a risk evaluation these devices may pose to end users and their patients. According to the Urgent Medical Device Recall Customer Notification Letter there are no harms associated with the use of affected reprocessed arthroscopic shavers. 1) Noncompatible devices are rendered inactive with console and alternative surgical methods are readily available. 2) Incorrect spin speed does not result in any harm to the patient The letter also informed customers to; 1) discontinue use of affected reprocessed arthroscopic shavers; 2) complete the recall effectiveness check form and indicate if any of the affected product remains in their inventory; 3)Return the completed form either via e-mail to SSSPFA@Stryker.com or to their Stryker Sustainability Solutions sales rep, or via fax to 480.763.5345; 4) If no product is in inventory customers are also being told to complete the recall effectiveness check form and send in as described above; 5) finally any adverse events or quality problems experienced with the use of this product should be reported to: The FDAs MedWatch Adverse Event Reporting program either online at https://www.accessdata.fda.gov/scripts/medwatch/index.cfm?action=reporting.home, by regular mail or by fax; Strykers Sustainability Solutions division complaint hotline: +1 888 763 8803 or http://www.stryker.com/productexperience
Initiated
2018-07-27
Root cause
Software Design Change
Terminated
2020-09-17
Distribution
Nationwide Distribution: AL, AR, AZ, CA, CO, DC, DE, FL, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY and Puerto Rico. OUS Israel and Canada.
Recall number
Z-0973-2019
Recall status
Terminated
CDRH record ID
170821
FDA record created
2019-03-05
Recall event number
82063

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗