Open original FDA record ↗Recall record
- FDA identifier
- Z-0975-2019
- Company in FDA record
- Draegar Medical Systems, Inc.
- FDA record creation date
- Mar 5, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm is recalling their Delta family of patient monitors software due to cybersecurity vulnerabilities, which may cause the device to reboot, lose functionality, and/or lose communication.
Source facts
- Action
- On February 8, 2019, the firm, Drager, notified affected customers of the recall via an "Urgent: Medical Device Recall" letter. Customers were informed that the existing software had cybersecurity vulnerabilities and that software version VF10.1 would mitigate these issues. Customers were asked to complete a Customer Acknowledgment and Response Form and return it to Drager via fax 215-372-2940 or email to ditelford.guality@draeger.com. Your local Draeger Service representative will contact you to schedule an appointment for upgrading your systems software once a compatible version has been released for distribution. If you are aware of any incidents related to this issue or if you have any further questions regarding operation of your patient monitor, please contact Drager Service Technical Support between the hours of 8:00 AM- 8:00 PM EST at 1-800-437-2437 (press 2 at the prompt, then 2 again).
- Initiated
- 2019-02-08
- Root cause
- Software design
- Distribution
- US Nationwide distribution including Puerto Rico.
- Recall number
- Z-0975-2019
- Recall status
- Open, Classified
- CDRH record ID
- 170565
- FDA record created
- 2019-03-05
- Recall event number
- 82014