Open original FDA record ↗Recall record
- FDA identifier
- Z-0976-04
- Company in FDA record
- Varian Medical Systems
- FDA classification date
- Jul 20, 2004
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Manufacturer of brachytherapy instrument issued a Customer Technical Bulletin to clarify instructions for use of the radiation treatment medical device by healthcare practitioners.
Source facts
- Action
- The recalling firm notified end users by letter and Customer Technical Bulletin on 5/4/04. The notification informed users not to use the Emergency retract hand wheel to recover a dummy wire and to reinforce the instructions already in the procedures that in the event of an Active wire becoming stuck out, the user should not turn the emergency retract hand wheel in excess of 20 times, or until the wire has parked and the radiation alarm has silenced. Advised of proper procedures for retracting the active wire and a warning concerning excessive turning of the emergency retract handwheel. The notification also states that customer user manuals and handwheel labeling will be updated at next service visit.
- Initiated
- 2004-05-04
- Root cause
- Other
- Terminated
- 2005-03-09
- Distribution
- The medical devices for correction had been distributed to medical facilities nationwide and internationally.
- Recall number
- Z-0976-04
- Recall status
- Terminated
- CDRH record ID
- 32958
- FDA classified
- 2004-07-20
- Recall event number
- 29005