Open original FDA record ↗Recall record
- FDA identifier
- Z-0976-2015
- Company in FDA record
- Becton Dickinson & Co.
- FDA classification date
- Jan 14, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A low level of surface and subsurface contamination of Listeria monocytogenes on 9 lots of non-sterile prepared plated media containing sheep blood, typically visible immediately upon removal from the packaging. The contamination is typically observed as 1-3 CFUs per plate when present.
Source facts
- Action
- BD Diagnostic Systems sent a Field Correction Notice dated August 2014 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to continue to follow all surveillance requirements of their laboratory upon confirmation of Listeria monocytogenes. Laboratory staff should notify their respective state health department. Customers with questions should contact BD Technical Services at 800-638-8663. For questions regarding this recall call 410-316-4000.
- Initiated
- 2014-06-06
- Root cause
- Material/Component Contamination
- Terminated
- 2016-08-12
- Distribution
- Worldwide Distribution - US nationwide including Puerto Rico, Belgium, Canada, Philippines, Singapore, and Taiwan.
- Recall number
- Z-0976-2015
- Recall status
- Terminated
- CDRH record ID
- 129615
- FDA classified
- 2015-01-14
- Recall event number
- 69091