Z-0977-2019: e600 Automatic Transport Ventilator, Model Number 01EVE600

Recall record Z-0977-2019 for O-Two Medical Technologies, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0977-2019
Company in FDA record
O-Two Medical Technologies, Inc.
FDA classification date
Apr 4, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.

Source facts

Action
Urgent Medical Device Correction notification letters dated 2/7/19 were sent to customers. An additional letter was sent to customers on 3/4/19.
Initiated
2019-02-12
Root cause
Process control
Terminated
2023-06-29
Distribution
The products were distributed to the following US states: AZ, CA, CO, FL, OH, PA, TX, and VA.
Recall number
Z-0977-2019
Recall status
Terminated
CDRH record ID
171093
FDA classified
2019-04-04
Recall event number
82178

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗