Open original FDA record ↗Recall record
- FDA identifier
- Z-0977-2024
- Company in FDA record
- Maquet Medical Systems USA
- FDA classification date
- Feb 2, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Nonconformities at their seal could potentially compromise the sterile barrier for affected products. Breach of sterile barrier could expose patients to pathogenic agents, resulting in the following potential harms: inflammation, infection, or sepsis.
Source facts
- Action
- On January 8, 2024, the firm notified affected customers via URGENT MEDICAL DEVICE Removal letters. Customers were instructed to remove all affected RF-32 pumps from use. If product is already in use, the firm advised to not discontinue therapy, but to monitor the patient closely for signs of infection. Customers should quarantine all affected product and return unopened/unexpired affected product to MCP/Getinge. Please contact MCP/Getinge Customer Support at (888) 943-8872 (press option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone) to request a return authorization (RMA) number and shipping instructions. Distributors should forward the recall document to customers who received affected products. Information also posted on firm's website: https://www.getinge.com/us/products-and-solutions/cardiovascular-procedures/ ***Updated 3/21/24*** Firm expanded field action to include additional lots. Amended notifications were sent to affected US customers on March 20, 2024.
- Initiated
- 2024-01-08
- Root cause
- Process control
- Distribution
- Worldwide - US Nationwide distribution to the states of CA, IL, MD, MO, NC, NY, OH, OR, PA, TX, UT, VA and the countries of Canada, Germany, Italy, Japan, and Netherlands.
- Recall number
- Z-0977-2024
- Recall status
- Open, Classified
- CDRH record ID
- 205668
- FDA classified
- 2024-02-02
- Recall event number
- 93852