The meter may self-start without a sample having been applied and report erroneous results without an error message if an undosed test strip is bent upwards while the meter is waiting for the sample or if the test strip opening has been previously flooded with excess control solution or blood.
Source facts
Action
A recall letter dated 5/18/05 was sent to health care professionals informing them of the problem and reqeusting them to provide a letter to their patients alerting them to the issue and how to avoid it. A notice of the problem will be placed in all packages of strips in the future.
Initiated
2005-05-18
Root cause
Other
Terminated
2006-02-26
Distribution
Nationwide.
Recall number
Z-0978-05
Recall status
Terminated
CDRH record ID
38940
FDA classified
2005-07-13
Recall event number
32021