A software anomaly in a medical device used for patient treatment could result in accidental or unintended exposure radiation exposure. The tilt rotation compensation algorithm does not adequately compensate for the tilt rotation.
Source facts
Action
The recalling firm notified medical device end users by letter dated 03/09/07 and advised of the software anomaly and provided details for user corrective action pending software upgrade by the manufacturer.
Initiated
2007-03-09
Root cause
Software design
Terminated
2012-02-28
Distribution
The medical devices were distributed to medical centers nationwide and worldwide.
Recall number
Z-0978-2007
Recall status
Terminated
CDRH record ID
51587
FDA classified
2007-06-28
Recall event number
37803