The inner diameter of the Sleeve shaft is undersized, resulting in interference fit when inserting the Anchored Lateral Straight Retractable Awl or the Anchored Lateral Straight Retractable Drill into the Sleeve shaft.
Source facts
Action
The firm, Lanx, sent their Consignees/customers a letter dated October 15, 2013. The letter described the product, problem and actions taken. Lanx and their distributors replaced the affected inventory with the conforming product.
Please contact the Chief Compliance Officer at 303-501-8534 or by email: Randy.fesmire@lanx.com.
Initiated
2013-08-22
Root cause
Component design/selection
Terminated
2015-03-13
Distribution
US Distribution to states of: AZ, CA, NY, TN, and TX including PR.
Recall number
Z-0978-2015
Recall status
Terminated
CDRH record ID
132196
FDA classified
2015-01-14
Recall event number
70060
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.