There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.
Source facts
Action
Urgent Medical Device Correction notification letters dated 2/7/19 were sent to customers.
An additional letter was sent to customers on 3/4/19.
Initiated
2019-02-12
Root cause
Process control
Terminated
2023-06-29
Distribution
The products were distributed to the following US states: AZ, CA, CO, FL, OH, PA, TX, and VA.
Recall number
Z-0978-2019
Recall status
Terminated
CDRH record ID
171094
FDA classified
2019-04-04
Recall event number
82178