Z-0978-2024: BO-RF-32-USA RotaFlow Centrifugal Pump with SOFTLINE Coating, Product Code 701047553. For use with an ext

Recall record Z-0978-2024 for Maquet Medical Systems USA. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0978-2024
Company in FDA record
Maquet Medical Systems USA
FDA classification date
Feb 2, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Nonconformities at their seal could potentially compromise the sterile barrier for affected products. Breach of sterile barrier could expose patients to pathogenic agents, resulting in the following potential harms: inflammation, infection, or sepsis.

Source facts

Action
On January 8, 2024, the firm notified affected customers via URGENT MEDICAL DEVICE Removal letters. Customers were instructed to remove all affected RF-32 pumps from use. If product is already in use, the firm advised to not discontinue therapy, but to monitor the patient closely for signs of infection. Customers should quarantine all affected product and return unopened/unexpired affected product to MCP/Getinge. Please contact MCP/Getinge Customer Support at (888) 943-8872 (press option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone) to request a return authorization (RMA) number and shipping instructions. Distributors should forward the recall document to customers who received affected products. Information also posted on firm's website: https://www.getinge.com/us/products-and-solutions/cardiovascular-procedures/ ***Updated 3/21/24*** Firm expanded field action to include additional lots. Amended notifications were sent to affected US customers on March 20, 2024.
Initiated
2024-01-08
Root cause
Process control
Distribution
Worldwide - US Nationwide distribution to the states of CA, IL, MD, MO, NC, NY, OH, OR, PA, TX, UT, VA and the countries of Canada, Germany, Italy, Japan, and Netherlands.
Recall number
Z-0978-2024
Recall status
Open, Classified
CDRH record ID
205669
FDA classified
2024-02-02
Recall event number
93852

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗