Skip to content

Public FDA record

Z-0979-2024: Barco MDSC-8527 NXF, Part Number K9352421

Recall record Z-0979-2024 for Barco N.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0979-2024
Company in FDA record
Barco N.V.
FDA classification date
Feb 2, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Barco became aware of a problem with I2C cables in the manufacturing process of a limited batch of MDSC-85xx NXF products. The functionality of the built-in MNA decoder is blocked.

Source facts

Action
Barco sent an email to their sole consignee on 12/14/2023 regarding the issue with the device and followed with a Customer Notice issued on 12/20/2023, also by email. The notice explained the issue, potential risk, and requested the following actions: A) Assess & locate the affected units and block them for further distribution. Please identify affected devices and find out where they are. These units should not be further processed within the distribution chain. B) Return the units to Barco Please return the products to Barco using the RMA form received by Barco. This form will be communicated in the next day. Please provide pick up details accordingly. C) Return Acknowledgement Letter to Barco upon completion of the actions. Please complete the Acknowledgement Letter provided in this document and send it back to your Barco contact after you have completed the actions described in this letter. The consignee was directed to ensure that your customers and users who have received affected products, are made aware of this notification and of the recommended actions and that you are the primary contact for them concerning this issue. Please provide each of your customers or users with a copy of this notice.
Initiated
2023-12-20
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the state of New Jersey.
Recall number
Z-0979-2024
Recall status
Open, Classified
CDRH record ID
205277
FDA classified
2024-02-02
Recall event number
93712

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗