Pump keyboard entries by the patient could have resulted in the patient having unintended access to programming screens and led to in improper drug dosage.
Source facts
Action
A recall letter, sent to the consignees on June 23, stated that the firm would telephone the consignees to arrange the correction of their affected pumps by reprogramming with the new software.
Initiated
2003-06-23
Root cause
Other
Terminated
2004-07-02
Distribution
The product was distributed only to two hospitals, one in North Carolina and one in Washington State.
Recall number
Z-0980-03
Recall status
Terminated
CDRH record ID
27876
FDA classified
2003-07-03
Recall event number
26550