Z-0980-05: LifeScan brand One Touch UltraSmart Meter Owners Booklet AW 060-788-01A; All One Touch UltraSmart Meters

Recall record Z-0980-05 for Lifescan Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0980-05
Company in FDA record
Lifescan Inc
FDA classification date
Jul 13, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Meter labeling (Owner's Booklet and Quick Reference Guide) is incorrect as it indicates that the meter displays the message 'Do you need a snack' for glucose values between 20 and 70 mg/dL; whereas the OneTouch UltraSmart meter displays the message, 'Do you need a snack?' for glucose values from 20 through 59 mg/dL.

Source facts

Action
On 6/9/05, the firm initiated the recall and their notification was via letters informing its customers of the corrective action.
Initiated
2005-06-09
Root cause
Other
Terminated
2006-12-21
Distribution
Product was released for distribution to 40,763 consignees worldwide. The firm provided the following distribution information: 124 wholesaler dealers, 224 distributors, 140 retailers and 40,275 consumers/users involved in the recall. The recall product was distributed in all States within the U.S. In addition, OneTouch UltraSmart meter kits with the affected UltraSmart Owner''s Booklets (AW 060-788-01A) have been distributed internationally in Canada, France, Germany, Italy, United Kingdom, Benelux, Sweden, Finland, Norway, Denmark, Spain, Portugal, Switzerland and Austria. The recall was appropriately extended to the consumer/user level; i.e., the wholesalers/distributors, pharmacists, and patients who received the recalled products.
Recall number
Z-0980-05
Recall status
Terminated
CDRH record ID
39887
FDA classified
2005-07-13
Recall event number
32360

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗