Open original FDA record ↗Recall record
- FDA identifier
- Z-0980-2007
- Company in FDA record
- Ethicon Endo-Surgery Inc
- FDA classification date
- Jun 28, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The adhesive used to bond the casing of the device may have migrated to the staple driver. While the device still fires, it may result in audible and tactile feedback at the midpoint of the firing cycle which may be interpreted by some surgeons as a completed firing cycle, which may cause the surgeon to prematurely stop firing the device and result in incomplete staple formation or cut line.
Source facts
- Action
- The recalling firm sent a recall letter dated 04/02/07 to foreign consignees and a second letter dated 04/03/07 to their United States consignees. The letter asks that the customers: 1) Examine their inventories for the presence of the affected product and remove it from inventory if found; 2) Complete the Business Reply Form and fax it to Stericycle their representative and 3) Return all affected product in order to received credit, along with the completed Business Reply Form in the box identified with a pre-printed UPS shipping label supplied by the recalling firm.
- Initiated
- 2007-04-02
- Root cause
- Other
- Terminated
- 2009-12-09
- Distribution
- Worldwide, including USA and Canada.
- Recall number
- Z-0980-2007
- Recall status
- Terminated
- CDRH record ID
- 52012
- FDA classified
- 2007-06-28
- Recall event number
- 37872