Skip to content

Public FDA record

Z-0981-05: Clinician Workstation Software, product code 2M5042

Recall record Z-0981-05 for Baxter Healthcare Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0981-05
Company in FDA record
Baxter Healthcare Corp.
FDA classification date
Jul 13, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The medication administration results displayed in the 'View Results' link may be inaccurate.

Source facts

Action
Urgent Device Correction recall letters dated 3/10/05 were sent to the direct accounts on the same date via first class mail. The letter describes the inaccuracies in the 'View Results' program, and informed the accounts that inaccurate information could lead to incorrect clinical decision-making, which could impact a patient''s medication treatment protocol. The accounts were requested to disable the 'View Results' link until a software upgrade is available, and to inform all clinical point of care users that the 'View Results' has been disabled and that the Manual Adminstration Record is to be used as the primary source of information for reviewing medication adminstration results. The accounts were instructed to contact their account executive and/or the Client Support Services at 1-877-835-2727 for instructions on how to disable the 'View Results' feature.
Initiated
2005-03-10
Root cause
Other
Terminated
2008-02-06
Distribution
Missouri, Massachusetts, Vermont, California and Canada.
Recall number
Z-0981-05
Recall status
Terminated
CDRH record ID
37948
FDA classified
2005-07-13
Recall event number
31467

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗