Open original FDA record ↗Recall record
- FDA identifier
- Z-0981-2019
- Company in FDA record
- Becton Dickinson & Co.
- FDA record creation date
- Mar 6, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
False negative results with the BD MAX Extended Enteric Bacterial Panel when testing specifically for Enterotoxigenic Escherichia coli (ETEC) and Plesiomonas shigelloides targets of the BD MAX Extended Enteric Bacterial Panel assay.
Source facts
- Action
- BD Diagnostic Systems issued "URGENT MEDICAL DEVICE PRODUCT ADVISORY" Customer letters via UPS on January 14, 2019. The letters described the product, problem, and actions to take. The customers were instructed to do the following: 1. Complete the attached Customer Response Form and return to BD Regulatory Compliance at Email: BDRC3@bd.com or Fax to: No.: 410-316-4258 whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification. 2. Although prevalence rates are considered to be low, if no other cause of enteric infection is determined you may choose to confirm any negative result for Enterotoxigenic Escherichia coli (ETEC) and Plesiomonas shigelloides obtained using the BD MAX Extended Enteric Bacterial Panel assay using an alternate method or using sample buffer tubes from a lot of EBP not listed above. Note that treatment for these infections is mainly supportive and antibiotics are not commonly used except in severe cases. 3. If your laboratory uses the xEBP assay and you would like replacement of your EBP kit(s), please complete the acknowledgement form included in the letter and return it via the instructions on the form. 4. Share this Advisory Letter with all users of the BD MAX instrument within your facility to ensure awareness. BD will provide replacement upon request using the Customer Response form. If you require further assistance, please contact: BD Customer/Technical Support 800-638-8663 Monday Friday between 7:00am and 7:00pm (EST) in the United States.
- Initiated
- 2019-01-14
- Root cause
- Nonconforming Material/Component
- Terminated
- 2020-04-03
- Distribution
- Worldwide Distribution: US (nationwide) to states of:CA, GA, HI, ID, MI, MN, MT, NJ, NM, OH, OR, PA, SC, TX, VA, VT, WA ; and to countries of: Australia, New Zealand, Canada, Europe, and Korea.
- Recall number
- Z-0981-2019
- Recall status
- Terminated
- CDRH record ID
- 170767
- FDA record created
- 2019-03-06
- Recall event number
- 82046