Z-0981-2019: BD MAX Enteric Bacterial Panel -For In Vitro Diagnostic Use with the BD MAX System Catalog Number: 442963

Recall record Z-0981-2019 for Becton Dickinson & Co.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0981-2019
Company in FDA record
Becton Dickinson & Co.
FDA record creation date
Mar 6, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

False negative results with the BD MAX Extended Enteric Bacterial Panel when testing specifically for Enterotoxigenic Escherichia coli (ETEC) and Plesiomonas shigelloides targets of the BD MAX Extended Enteric Bacterial Panel assay.

Source facts

Action
BD Diagnostic Systems issued "URGENT MEDICAL DEVICE PRODUCT ADVISORY" Customer letters via UPS on January 14, 2019. The letters described the product, problem, and actions to take. The customers were instructed to do the following: 1. Complete the attached Customer Response Form and return to BD Regulatory Compliance at Email: BDRC3@bd.com or Fax to: No.: 410-316-4258 whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification. 2. Although prevalence rates are considered to be low, if no other cause of enteric infection is determined you may choose to confirm any negative result for Enterotoxigenic Escherichia coli (ETEC) and Plesiomonas shigelloides obtained using the BD MAX Extended Enteric Bacterial Panel assay using an alternate method or using sample buffer tubes from a lot of EBP not listed above. Note that treatment for these infections is mainly supportive and antibiotics are not commonly used except in severe cases. 3. If your laboratory uses the xEBP assay and you would like replacement of your EBP kit(s), please complete the acknowledgement form included in the letter and return it via the instructions on the form. 4. Share this Advisory Letter with all users of the BD MAX instrument within your facility to ensure awareness. BD will provide replacement upon request using the Customer Response form. If you require further assistance, please contact: BD Customer/Technical Support 800-638-8663 Monday Friday between 7:00am and 7:00pm (EST) in the United States.
Initiated
2019-01-14
Root cause
Nonconforming Material/Component
Terminated
2020-04-03
Distribution
Worldwide Distribution: US (nationwide) to states of:CA, GA, HI, ID, MI, MN, MT, NJ, NM, OH, OR, PA, SC, TX, VA, VT, WA ; and to countries of: Australia, New Zealand, Canada, Europe, and Korea.
Recall number
Z-0981-2019
Recall status
Terminated
CDRH record ID
170767
FDA record created
2019-03-06
Recall event number
82046

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗