Skip to content

Public FDA record

Z-0982-2015: Anchored Lateral Retractable Drill, Timberline IPF IFU IFU8621-0113, Model 8630-0102. Timberline Anchor

Recall record Z-0982-2015 for Biomet Spine, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0982-2015
Company in FDA record
Biomet Spine, LLC
FDA classification date
Jan 15, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The outer diameter of the Drill shaft is oversized; thereby, resulting in interference fit when inserting the Drill into the Fixed or Variable Sleeve Assembly.

Source facts

Action
Distributors and Direct Sales Reps notified via letter to return product.
Initiated
2013-08-22
Root cause
Component design/selection
Terminated
2015-03-13
Distribution
Distributed in PR and the states of NY, AZ, TN, TX, and CA.
Recall number
Z-0982-2015
Recall status
Terminated
CDRH record ID
132193
FDA classified
2015-01-15
Recall event number
70056

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗