The inner diameter of the Sleeve shaft is undersized, resulting in interference fit when inserting the Anchored Lateral Straight Retractable Awl or the Anchored Lateral Straight Retractable Drill into the Sleeve.
Source facts
Action
The firm, Lanx, sent a letter to their Consignees/customers dated October 15, 2013. The letter describes the product, problem and actions taken. Lanx and their distributors replaced the affected inventory with conforming product.
Please contact Chief Compliance Officer at 303-501-8534 or by email: Randy.fesmire@lanx.com if you have any questions regarding this recall.
Initiated
2013-08-22
Root cause
Device Design
Terminated
2015-03-13
Distribution
US Distribution to states of: AZ, CA, NY, TN and TX including PR.
Recall number
Z-0983-2015
Recall status
Terminated
CDRH record ID
132207
FDA classified
2015-01-15
Recall event number
70070
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.