Open original FDA record ↗Recall record
- FDA identifier
- Z-0983-2017
- Company in FDA record
- Bausch & Lomb, Inc.
- FDA classification date
- Jan 10, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
According to firm ( Bausch & Lomb, Inc. ) on June 17, 2016 a packaging operator notified his supervisor he had not placed the operating instruction into the last repackaging of SKU 0816 C (Pterygium Burr Medium Diamond). Further investigation found that repackaged Algerbrush-II and its accessories from April 30, 2012 to present, contained out of date operating instruction. Therefore Bausch & Lomb is proposing a Field Correction to provide customer purchasing product from April 30, 2012 to June 20, 2016 with the current operating instruction.
Source facts
- Action
- Bausch + Lomb sent an Urgent Medical Device Voluntary Field Correction letter dated October 12, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. For questions contact Bausch + Lomb customer service at 1-800-338-2020
- Initiated
- 2016-10-12
- Root cause
- Error in labeling
- Terminated
- 2017-07-14
- Distribution
- Worldwide Distribution - US Nationwide
- Recall number
- Z-0983-2017
- Recall status
- Terminated
- CDRH record ID
- 151706
- FDA classified
- 2017-01-10
- Recall event number
- 75915