Open original FDA record ↗Recall record
- FDA identifier
- Z-0983-2019
- Company in FDA record
- Maquet Cardiovascular, LLC
- FDA record creation date
- Mar 7, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential lack of carbon dioxide (CO2) flow that may result in a procedural delay during off-pump coronary artery bypass (OPCAB).
Source facts
- Action
- The firm, Maquet/Getinge, issued "URGENT MEDICAL DEVICE RECALL - REMOVAL" letters mailed to customers on January 24, 2019 via FedEx Priority Overnight,. The letters identified the issue, the product, actions to take and health risk. The customers were instructed to:Please examine your inventory immediately to determine if you have any of the affected; stop using the affected lot numbers of the Axius Blower Mister; remove and properly label any identified affected product, and keep it in quarantine in a secure location to preclude being used. Should you have un-used affected product you are eligible for either a replacement or credit. Please contact Geting Customer Service at 1-888-627-8383 (press option 2, then option 4) to request a return authorization (RMA) and shipping instructions to return any affected product. Complete and sign the MEDICAL DEVICE RECALL - REMOVAL RESPONSE FORM to acknowledge that you have received this notification. Return the completed form to Maquet/Getinge by e-mailing a scanned copy to blowermister(@qetinge.com or by faxing the form to 1-800-915-0124. If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. If you have any questions, please contact your Getinge representative or call the Getinge Customer Support at 1-888-627-8383 (press option 2, then option 4), Monday through Friday from 6:00 a.m. and 5:00 p.m. PST.
- Initiated
- 2019-01-24
- Root cause
- Employee error
- Terminated
- 2020-07-13
- Distribution
- Worldwide distribution: US (Nationwide) and countries of: Australia, Belgium, Canada, Czech Republic, France, Germany, Korea, Netherlands, Russia, Spain, Switzerland, and United Kingdom.
- Recall number
- Z-0983-2019
- Recall status
- Terminated
- CDRH record ID
- 170889
- FDA record created
- 2019-03-07
- Recall event number
- 82090