Skip to content

Public FDA record

Z-0983-2024: Dual Articulating Headrest, Pad Ret P/N P141210805, REF BF754

Recall record Z-0983-2024 for Steris Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0983-2024
Company in FDA record
Steris Corporation
FDA classification date
Feb 5, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Dual-articulating head rest may not stay in place or may fail to lock into position during a patient procedure which could result in an injury to the patient or a procedural delay.

Source facts

Action
An Urgent Medical Device Recall notice dated December 5, 2023 was issued to distributors and customers via FedEx explaining the removal of specific affected product . The notice explains the labeled date of manufacture must fall between August 14, 2023 and October 26, 2023 the recall notice must have been received directly from STERIS. The Dual-Articulating Head Rest may not stay in place or may fail to lock into position during a patient procedure which could result in an injury to the patient or a procedural delay. STERIS Service Technician will work with your facility to schedule a time to hand-deliver replacement head rest(s) and pick up your affected head rest(s). It is recommended all head rests impacted by this recall be immediately removed from service until it can be replaced. If you have further distributed this product, please identify all locations, notify them of this recall, and coordinate the collection of all affected head rest(s).
Initiated
2023-12-12
Root cause
Process control
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, FL, IL, IN, IA, KS, LA, MN, MS, NV, NJ, OH, OK, OR, PA, TN, TX, UT, WA, WI & Puerto Rico, and the countries of Australia, Canada, Mexico & New Zealand.
Recall number
Z-0983-2024
Recall status
Open, Classified
CDRH record ID
205326
FDA classified
2024-02-05
Recall event number
93733

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗