Philips Medical Systems North America Co. Phillips
FDA classification date
Jun 27, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
Potential for gaseous helium to improperly vent, resulting in rupture of the magnet vessel and venting of helium into the MRI site.
Source facts
Action
On 4/30/07 Philips sent an Urgent Device Correction letter dated April 25, 2007 via certified mail to customers. The letter advises customers of the issue and that that a representative will visit the sites, inspect and correct the problem if indicated.
Initiated
2007-04-30
Root cause
Other
Terminated
2011-09-21
Distribution
Units were distributed throughout the US to 1,078 hospitals and medical centers.
Recall number
Z-0984-2007
Recall status
Terminated
CDRH record ID
52335
FDA classified
2007-06-27
Recall event number
37950