Open original FDA record ↗Recall record
- FDA identifier
- Z-0984-2015
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Mar 18, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There may be an existing dark current phenomenon on ARTISTE LINAC in combination with IMRT or mARC treatments using unflat beams. Software issue.
Source facts
- Action
- Siemens Planned Action: 1. Providing recommendations for user actions in terms of routine dosimetric quality control checks for any IMRT treatment at very high dose rates (Multiple X) as well as for mARC treatments. 2. Software update for the Function Controller #3, which controls parts of the beam generation. This software update enables the correct switching between Pulse Forming Network (PFN) and reduced Pulse Forming Network (iPFN) in pause state, to reduce the dark current during the movement as much as possible. This software will be installed by a Siemens service representative. 3. Siemens will contact the user to make an appointment to implement the modification system at a date and time when the system will be available. These corrections shall be made free of charge. The deployment of the permanent correction shall be completed by July 1, 2015. For further questions please call (610) 219-6300.
- Initiated
- 2014-11-19
- Root cause
- Software design
- Terminated
- 2015-04-28
- Distribution
- US Distribution to the states of UT, WI, and NY.
- Recall number
- Z-0984-2015
- Recall status
- Terminated
- CDRH record ID
- 131894
- FDA classified
- 2015-03-18
- Recall event number
- 69880