Open original FDA record ↗Recall record
- FDA identifier
- Z-0984-2019
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA record creation date
- Mar 7, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The swivel arm holding the monitors may become dislodged from the carrying arm.
Source facts
- Action
- The firm, Siemens, issued an "URGENT: MEDICAL DEVICE CORRECTION "on 2/8/19 to its customers. The letter is identifying the problem, risk, actions to take and corrective action. The customers were instructed to do the following: Immediately examine your inventory and quarantine product subject to recall (if applicable). or To avoid the risk until implementation of the corrective action, it is strongly recommended not to move and position monitors above the system table. As an alternative and if clinically applicable, monitors can also be used when they are left in the parking position. In addition, if you may have further distributed this product, please identify your customers and notify them at once of this product recall. If the device has been sold and is therefore no longer in your possession, please forward this notice to the new owner. The service organization will contact and arrange a date to perform this corrective action with Updated Instruction XP006/19/S. Contact the service organization for an earlier appointment at 1-800-888-7436.
- Initiated
- 2019-02-08
- Root cause
- Device Design
- Terminated
- 2020-05-05
- Distribution
- US Nationwide Distribution
- Recall number
- Z-0984-2019
- Recall status
- Terminated
- CDRH record ID
- 171135
- FDA record created
- 2019-03-07
- Recall event number
- 82211