Open original FDA record ↗Recall record
- FDA identifier
- Z-0985-2009
- Company in FDA record
- Zimmer Inc.
- FDA classification date
- Mar 2, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Incorrect parameters: If the user switches from one surgeon profile to another surgeon profile after completing anatomical landmark acquisition tasks, the application will continue to use the initial surgeon profile parameter values in the calculation of the predicted hip-knee-ankle alignment estimates, which may have different resection or component thickness parameters than are appropriate for
Source facts
- Action
- Zimmer notified consignees via letter dated 8/19/08, which advised them not to change the selected surgeon profile after the acquisition of landmarks has begun, and the issued will be fixed with the next software release.
- Initiated
- 2008-08-19
- Root cause
- Software design
- Terminated
- 2010-04-28
- Distribution
- Worldwide distribution: USA, Australia, Austria, Belgium, Canada, France, Germany, Italy, Singapore, South Africa, Switzerland and United Kingdom.
- Recall number
- Z-0985-2009
- Recall status
- Terminated
- CDRH record ID
- 76911
- FDA classified
- 2009-03-02
- Recall event number
- 50746