Open original FDA record ↗Recall record
- FDA identifier
- Z-0985-2012
- Company in FDA record
- Philips Healthcare Inc.
- FDA classification date
- Feb 8, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Incorrect Skin Dose calculation. AKR (Air Kerma Rate) displayed and stored in patient records is too low
Source facts
- Action
- Philips Healthcare issued a mandatory Urgent Field Safety Noticedated October 3, 2011 to all affected customers. The letter identifies the affected product, problem description, hazard involved, action to be taken by customer/user and actions planned by Philips to correct the problem. The letter provide important information for the continued safe and proper use of the equipment. The letter instructs customers to review the following information with all members of their staff who need to be aware of the contents of this communication and retain a copy with the equipment Instruction for Use. For further information or support concerning this issue contact your local Philips representative at 866-767-2822.
- Initiated
- 2012-01-09
- Root cause
- Employee error
- Terminated
- 2017-02-13
- Distribution
- Worldwide Distribution - USA (nationwide) including the states of CA, IL, OH, SD, TX, WY and the countries of: Australia, Austria, Canada, Egypt, Germany, mexico, Turkey and UK
- Recall number
- Z-0985-2012
- Recall status
- Terminated
- CDRH record ID
- 106736
- FDA classified
- 2012-02-08
- Recall event number
- 60898