Skip to content

Public FDA record

Z-0986-2019: Cardiohelp Emergency Drive, Part No. 70104.8002 Product Usage: The Cardiohelp Emergency Drive is used in

Recall record Z-0986-2019 for Maquet Cardiovascular Us Sales, Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0986-2019
Company in FDA record
Maquet Cardiovascular Us Sales, Llc
FDA classification date
Mar 28, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Upon servicing the gearwheel in the Emergency Drive the gearwheel may have been re-assembled incorrectly, leading to the inadequate operation of the Emergency Drive.

Source facts

Action
Maquet has issued an Urgent Medical Device Recall dated February 26, 2019 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were informed that their Emergency Drives had already been inspected and replaced with a loaner Emergency Drive if found to be affected. Affected E-Drives were returned to Getinge for servicing. Upon servicing, a service technician will return the Emergency Drive to your facility. Customers were asked to complete and sign the customer response form to acknowledge receipt of the notification. If you have any questions, please contact your Maquet Cardiopulmonary (MCP)/Getinge representative or call the Maquet Cardiopulmonary (MCP)/Getinge Customer Support at +1 (888) 627-8383 (press option 2, then option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
Initiated
2019-02-08
Root cause
Process control
Terminated
2020-05-14
Distribution
US Nationwide Distribution in the states of California, Florida, Illinois, Kentucky, Michigan, and New York.
Recall number
Z-0986-2019
Recall status
Terminated
CDRH record ID
171237
FDA classified
2019-03-28
Recall event number
82265

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗