Z-0986-2020: Cocoon Convective Warming System, Product Code: CWS4000 (110V).

Recall record Z-0986-2020 for Care Essentials Pty., Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0986-2020
Company in FDA record
Care Essentials Pty., Ltd.
FDA record creation date
Mar 3, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for thermal damage near the power cord connector and Power Entry Module (PEM) due to improper connection of the power cord connector and PEM pins on the Cocoon Convective Warming System.

Source facts

Action
On January 7, 2020 the firm distributed Urgent Medical Device Recall Communication letters by FedEx stating: URGENT- Medical Device Field Correction Dear Valued Customer: Care Essentials (CE) is writing to advise you that, as a precautionary measure, we are issuing a voluntary product repair of CWS4000 and CWS5000 (110V) Cocoon Convective Warming Machines devices due to a report of thermal damage to the power entry module and power cord in a hospital in Canada. The affected devices were distributed in the U.S. from the date of March 3rd, 2015. There are no reported instances of death or serious injuries. However, in order to prevent a potential risk of fire, we are undertaking the following actions: - attaching a P-clip to the power cord of the device to both reduce the risk of using power cords that are not supplied or approved by Care Essentials, as well as reducing wear and tear of the power cord connector contacts from repeated plugging and unplugging; and - updating the instructions for use (IFU") to add warnings about avoiding unnecessary manipulation of the power cord connector, using only power cords (which have P-clips attached) supplied by Care Essentials, and not properly using a Care Essentials P-clipped power cord could lead to risk of fire. We ask that you: - Inspect devices cited in this notification letter for any damage to the power cord connector and power socket. If any damage is observed to these parts, please do not use the affected device and contact the undersigned for repair or replacement. - Ensure that this communication, including the updated IFU with highlighted revisions, is shared with all departments within the hospital/group of hospitals within your organization that may be using one of these devices. - Confirm receipt of this letter by email (msok@msi-healthcare.com ), or queries@careessentials.com.au) or Telephone (703-9957179). Kindly fill and return the attached form: Medical Device Field Correction R
Initiated
2020-01-07
Root cause
Other
Terminated
2023-03-22
Distribution
US Nationwide in the states of AZ CA FL GA KS LA MA MD MI MN NC NM NY OH PA QC SC TX VA WV and the Foreign distribution to Canada.
Recall number
Z-0986-2020
Recall status
Terminated
CDRH record ID
178562
FDA record created
2020-03-03
Recall event number
84541

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗