Enabling sending of preliminary/initial results on the ADVIA Chemistry XPT creates a risk that a critical result could be overwritten by the same result rather than showing the repeated result on CentraLink. A result could be erroneous and critical and appear to be verified upon repeat.
Source facts
Action
A Siemens service provider visited each site to verify and/or update customer configuration on the ADVIA Chemistry XPT and CentraLink Systems to ensure proper reporting of results. An Urgent Medical Device Correction letter was issued on December 16, 2014 to all affected customers. Customers were also provided a Customer Bulletin providing instructions on how configuration setting should remain.
Initiated
2014-12-16
Root cause
Under Investigation by firm
Terminated
2016-10-27
Distribution
Distributed in the state of WA.
Recall number
Z-0987-2015
Recall status
Terminated
CDRH record ID
132219
FDA classified
2015-01-15
Recall event number
70078