Z-0987-2019: Maquet XS Flat Screen Monitor Holder, XS32 SPE, Configuration Part Number ARD22201040A Product Usage: Th

Recall record Z-0987-2019 for GETINGE US SALES LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0987-2019
Company in FDA record
GETINGE US SALES LLC
FDA classification date
Mar 7, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The affected flat screen monitor holders might not be compliant with grounding requirements. As a consequence, there is a risk of mild to severe/fatal electric shock under certain conditions while using the flat screen monitor holder with voltage of 100-220V.

Source facts

Action
Urgent Medical Device Field Correction notification letters dated 1/8/19 were sent to customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to: -Complete and sign the enclosed Customer Response Form (Pages 4 and 5) and return to Getinge using the fax number or email address provided on the form. -A Getinge Service person will contact the person listed on the Customer Response Form (page 5) to schedule service for an update of your device, free of charge. You can also arrange a visit by a Getinge-trained or authorized service technician by calling Getinge Customer Support at (800) 950-9912 (select option 1).
Initiated
2019-01-08
Root cause
Other
Terminated
2020-06-03
Distribution
The products were distributed to the following US states: AZ and TX.
Recall number
Z-0987-2019
Recall status
Terminated
CDRH record ID
170848
FDA classified
2019-03-07
Recall event number
82074

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗