Z-0987-2025: Merlin.net Patient Care Network (PCN) Heart Failure Web Application (Model MN6000) (Which interfaces with

Recall record Z-0987-2025 for St. Jude Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0987-2025
Company in FDA record
St. Jude Medical
FDA classification date
Jan 17, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a data migration from on-premises server to the cloud environment, some customers reported that they were not able to transmit readings.

Source facts

Action
On 11/06/2024, the firm sent an "URGENT: MEDICAL DEVICES CORRECTION", to customers to inform them that on September 30, 2024, the Merlin.net PCN HF Portal (MN6000) was migrated to the cloud database and following the migration, some customers reported that their Hospital Systems were unable to transmit to the Merlin.net PCN. User Action Requested: Abbott requests that clinicians take the following actions: The Impacted CM3100 Hospital System devices at your facility is/are listed in the enclosed Appendix. o For a Hospital System marked SESSIONS SENT no further correction is needed. o Hospital Systems marked UNSENT SESSIONS should be turned on to send unsent sessions: If your Hospital System has connectivity, it will immediately attempt to send after boot-up. A spinning icon will appear to indicate that transmission is in progress. If you see a button on the home screen showing X Unsent Sessions select the button and follow the prompts to send. If you are unable to send sessions, contact Abbott Remote Care Technical Support (RCTS) for assistance. Distribute this notice to appropriate personnel within your institution. Complete the included acknowledgment form and return it to Abbott. NOTE: Hospital Systems that were not used between September 30, 2024, and October 15, 2024, are not impacted. any questions about this notice or potential patient impact, please contact your Abbott representative or Remote Care Technical Support at 1-844-MYCMEMS (692-6367) Monday Friday, 5AM PST 5PM PST.
Initiated
2024-11-06
Root cause
Under Investigation by firm
Distribution
U.S.: CA, CT, GA, IL, MI, MN, NC, NJ, NY, OR, PA, TX, and WA. O.U.S.: N/A
Recall number
Z-0987-2025
Recall status
Open, Classified
CDRH record ID
210981
FDA classified
2025-01-17
Recall event number
95706

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗