Open original FDA record ↗Recall record
- FDA identifier
- Z-0988-2019
- Company in FDA record
- GETINGE US SALES LLC
- FDA classification date
- Mar 7, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The affected flat screen monitor holders might not be compliant with grounding requirements. As a consequence, there is a risk of mild to severe/fatal electric shock under certain conditions while using the flat screen monitor holder with voltage of 100-220V.
Source facts
- Action
- Urgent Medical Device Field Correction notification letters dated 1/8/19 were sent to customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to: -Complete and sign the enclosed Customer Response Form (Pages 4 and 5) and return to Getinge using the fax number or email address provided on the form. -A Getinge Service person will contact the person listed on the Customer Response Form (page 5) to schedule service for an update of your device, free of charge. You can also arrange a visit by a Getinge-trained or authorized service technician by calling Getinge Customer Support at (800) 950-9912 (select option 1).
- Initiated
- 2019-01-08
- Root cause
- Other
- Terminated
- 2020-06-03
- Distribution
- The products were distributed to the following US states: AZ and TX.
- Recall number
- Z-0988-2019
- Recall status
- Terminated
- CDRH record ID
- 170849
- FDA classified
- 2019-03-07
- Recall event number
- 82074