Open original FDA record ↗Recall record
- FDA identifier
- Z-0988-2020
- Company in FDA record
- Zimmer Biomet, Inc.
- FDA record creation date
- Feb 4, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for the expiration date on the outer label that was applied by Zimmer Biomet to exceed the expiration date on the inner label
Source facts
- Action
- Zimmer Biomet issued Urgent Device Recall dated 12/23/19, to Distributors/Hospitals via email/courier stating reason for recall, health risk and action to take: Distributors responsibilities- locating and removing the product in their territory, as well as identifying hospitals and surgeons who have previously used the product. Distributors will return on-hand product to Zimmer Biomet and ensure all of their products are accounted for using the form provided in the letter. Hopsitals: Review this notification and ensure that affected personnel are aware of the contents, including surgeons that utilized this product. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. Questions or concerns call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday.. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com
- Initiated
- 2019-12-23
- Root cause
- Packaging process control
- Terminated
- 2020-09-10
- Distribution
- US Nationwide
- Recall number
- Z-0988-2020
- Recall status
- Terminated
- CDRH record ID
- 179000
- FDA record created
- 2020-02-04
- Recall event number
- 84679