Skip to content

Public FDA record

Z-0989-04: Micromix Compounder Implementation Guide; this guide is distributed by Professional Services as part of t

Recall record Z-0989-04 for Baxter Healthcare Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0989-04
Company in FDA record
Baxter Healthcare Corp.
FDA classification date
Jul 20, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The promotional material for the Automix/Accusource and Micromix compounders and Micromix Compounder Implementation Guide contain an incomplete accuracy statement regarding the delivery of each ingredient by the compounders.

Source facts

Action
Baxter sent Important Product Information letters dated 4/30/04 to all of the compounder customers, informing them of the inaccurate information and requesting that the Nutrition Compounding Simplified brochure, Baxter Info Quick Cards on CD-Rom, versions Q4 2003 and Q2 2003, and pages 60 and 61 of the Micromix Compounder Implementation Guide be destroyed. Replacement pages 60 and 61 of the implementation guide were enclosed with the letter. The letters also listed the accuracy information for each compounder based on Baxter''s supporting data, as well as the specific information regarding the overdelivery alarm limits referenced in the Operator''s Manuals.
Initiated
2004-04-30
Root cause
Other
Terminated
2006-04-21
Distribution
Nationwide, United Kingdom, Germany, Chile, Canada, Brazil, El Salvador, Taiwan, Korea, Italy, Saudi Arabia, Israel, China, Indonesia, Hong Kong, Colombia, Argentina, Mexico, Guatemala, Costa Rica, Panama and Honduras.
Recall number
Z-0989-04
Recall status
Terminated
CDRH record ID
32981
FDA classified
2004-07-20
Recall event number
29017

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗