- Z-0990-2017
- Company in FDA record
- Varian Medical Systems, Inc.
- FDA classification date
- Jan 10, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Modifications in version 13.6MR2 for Contouring, SmartAdapt, and SmartSegmentation workspaces resulted in contours not being saved consistently in Eclipse. Treatment Planning System. The issue only occurs if certain conditions are fulfilled.
Source facts
- Action
- Varian Medical Systems sent an Urgent Medical Device Correction Letter dated December 5, 2016 , that were sent to all 5 affected sites on December 5, 2016 by a trackable method. Letters advised users of the issue and the specific scenario that must take place for the issue to occur. Varian is investigating the issue and will provide a correction once it is available. We sincerely apologize for any inconvenience and thank you in advance for your cooperation. If you require further clarification, please feel free to contact your local Varian Medical Systems Customer Support District or Regional Manager. This notice has been provided to the appropriate Regulatory Authorities. For further questions please call (650) 424-6833. Varian Oncology Help Desk Contact Information: Phone: +1.888.827.4265 Email: support @varian.com
- Initiated
- 2016-12-02
- Root cause
- Device Design
- Terminated
- 2017-05-12
- Distribution
- US Distribution to the states of : NJ, TN. OR and FL.
- Recall number
- Z-0990-2017
- Recall status
- Terminated
- CDRH record ID
- 151945
- FDA classified
- 2017-01-10
- Recall event number
- 76018