St. Jude Medical, Cardiac Rhythm Management Division
FDA record creation date
Apr 21, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II
Reason stated in the source
Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to
enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the inability to remotely monitor these devices and resulted in reduced battery longevity.
Source facts
Action
Abbott issued a Safety Notification For a subset of Gallant ICDS and CRT-D, Models CDVRA500Q, CDDRA500Q and CDHFA500Q on March 10, 2022. Notification was hand delivered in US.
The notification includes the Patient Management Recommendations:
-As the devices are unable to be followed using remote monitoring via Bluetooth connectivity, follow-ups should be conducted in-clinic using inductive telemetry.
- Abbott has developed a process to attempt recovery of Bluetooth communication via a device firmware upgrade. If remote monitoring via Bluetooth communication is desired, please work with your local Abbott representative to schedule a firmware upgrade.
Please complete and return the included Acknowledgement Form to CRMNotification@abbott.com and maintain a record of this notice to ensure
effectiveness of the communication.
For any question, please contact Abbott Technical Support at 1-800-722-3774 (U.S.).
Initiated
2022-03-10
Root cause
Component design/selection
Distribution
Worldwide distribution - US Nationwide distribution in the states of FL, KY and the country of France, Czech Republic.
Recall number
Z-0990-2022
Recall status
Open, Classified
CDRH record ID
192641
FDA record created
2022-04-21
Recall event number
89825