Z-0991-2015: Infinite 200 The Tecan Infinite 200 is a multifunctional microplate reader with injector option. The Infi

Recall record Z-0991-2015 for Tecan US, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0991-2015
Company in FDA record
Tecan US, Inc.
FDA classification date
Jan 15, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Incorrect lumi firmware version installed (E.027 instead of V2.00)

Source facts

Action
Tecan sent an Urgent Field Corrective Action letter dated October2014, to all affected consignees by via FedEx on/or about November 10, 2014. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed not to use te kinetic luminescence measurement workflow requiring movement of th eplate transport out during kinectice measurements until the firmware has been upgraded to current version 2.0. A Tecan service representative will contact you to arrange a time to install the firmware. Additionally, they may want to review prior measurement results obtained with this workflow. Customers with questions were advised to contacte their local Tecan Helpdesk. For questions regarding this recall call 919-361-5200, ext 19524.
Initiated
2014-11-10
Root cause
Employee error
Terminated
2015-12-21
Distribution
Worldwide Distribution - USA including MD and Internationally to Hong Kong, South Korea and the UK.
Recall number
Z-0991-2015
Recall status
Terminated
CDRH record ID
131449
FDA classified
2015-01-15
Recall event number
69739

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗