iView report displayed femoral cut measurements in the incorrect unit of measure (inches vs. mm).
Source facts
Action
ConForMIS intiated the recall by telephone on October 31, 2016, to the Sales Rep to request return of the kit.
For questions regarding this recall call 781-345-9001.
Initiated
2016-10-31
Root cause
Software Design Change
Terminated
2017-01-11
Distribution
Distribution to TX only
Recall number
Z-0991-2017
Recall status
Terminated
CDRH record ID
151856
FDA classified
2017-01-10
Recall event number
75969