Open original FDA record ↗Recall record
- FDA identifier
- Z-0992-2015
- Company in FDA record
- Tecan US, Inc.
- FDA classification date
- Jan 15, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
Incorrect lumi firmware version installed (E.027 instead of V2.00)
Source facts
- Action
- Tecan sent an Urgent Field Corrective Action letter dated October2014, to all affected consignees by via FedEx on/or about November 10, 2014. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed not to use te kinetic luminescence measurement workflow requiring movement of th eplate transport out during kinectice measurements until the firmware has been upgraded to current version 2.0. A Tecan service representative will contact you to arrange a time to install the firmware. Additionally, they may want to review prior measurement results obtained with this workflow. Customers with questions were advised to contacte their local Tecan Helpdesk. For questions regarding this recall call 919-361-5200, ext 19524.
- Initiated
- 2014-11-10
- Root cause
- Employee error
- Terminated
- 2015-12-21
- Distribution
- Worldwide Distribution - USA including MD and Internationally to Hong Kong, South Korea and the UK.
- Recall number
- Z-0992-2015
- Recall status
- Terminated
- CDRH record ID
- 131450
- FDA classified
- 2015-01-15
- Recall event number
- 69739