Open original FDA record ↗Recall record
- FDA identifier
- Z-0992-2019
- Company in FDA record
- Ortho-Clinical Diagnostics
- FDA record creation date
- Mar 9, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Imprecision in the affected lot may produce negatively or positively biased patient sample and quality control fluid results.
Source facts
- Action
- The firm, Ortho Clinical Diagnostics, sent an "URGENT PRODUCT CORRECTION NOTIFICATION" letters dated 1/22/19 to is customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: " If you have an alternate lot of VITROS Na+ Slides, discontinue using and discard all remaining inventory of Lot 4221-0998-3223. " If you do NOT have an alternate lot, consult with your medical or laboratory director to determine if it is acceptable to continue using the affected lot until your replacement order arrives. " Until you receive your replacement order: o Performing Osmolality calculations on your VITROS System is not recommended. o If you continue using the affected lot, it is strongly recommended to take additional action to detect imprecision, such as performing duplicate testing for each sample. Repeat testing if there is a discrepancy between the results obtained in duplicate testing (i.e., +/- 1.4 mmol/L). o Discard all remaining inventory when your replacement order arrives. An adequate supply of replacement product is available and Ortho will expedite all replacement orders. " Please complete the enclosed Confirmation of Receipt form by e-Mail to RA-OCDUS-CONFIRMAD@ORTHOCLINICALDIAGNOSTICS.COM or Fax to: 1.888.557.3759 or 1.585.453.4110. Ortho will credit your account or send a replacement order for the discarded slides. " Forward the information in this notification if you have distributed this product outside of your facility. If you have questions, please contact Ortho Care" Technical Solutions Center at 1-800-421-3311.
- Initiated
- 2019-01-22
- Root cause
- Under Investigation by firm
- Terminated
- 2022-06-10
- Distribution
- US Distribution to states of: AL, CA, CT, FL, HI, IA, ID, IL, LA, MI, MO, MS, NC, NM, NY, OK, PA, TX, WV, and WY; and internationally to: Canada.
- Recall number
- Z-0992-2019
- Recall status
- Terminated
- CDRH record ID
- 170972
- FDA record created
- 2019-03-09
- Recall event number
- 82113