Z-0992-2024: RX Series Copper (Cu) Assay Ref. Number CU2340

Recall record Z-0992-2024 for Randox Laboratories Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0992-2024
Company in FDA record
Randox Laboratories Ltd.
FDA classification date
Feb 6, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Randox Laboratories have released an update to the Carryover Avoidance Technical Bulletin to detail that Copper, should not be run directly after Total Protein, on the RX series instruments testing order. If copper assay is run after total protein it could lead to falsely elevated copper results.

Source facts

Action
RANDOX Distributors issued Medical Device Correction Letter (REC704) on December 12, 2023 to US consignees. Letter states reason for recall, health risk and action to take: Review your instrument testing order in line with the Carryover Avoidance Technical Bulletin (RXTB-0148) " Update the RX user manual with the updated Carryover Avoidance Technical Bulletin (RXTB-0148) and ensure all operators are aware of the recommendations. " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organization.
Initiated
2023-11-29
Root cause
Component design/selection
Distribution
US Nationwide distribution in the states of SC and OK.
Recall number
Z-0992-2024
Recall status
Open, Classified
CDRH record ID
205416
FDA classified
2024-02-06
Recall event number
93775

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗