Open original FDA record ↗Recall record
- FDA identifier
- Z-0993-2011
- Company in FDA record
- Arstasis, Inc
- FDA classification date
- Feb 4, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Device has the potential to fracture and sheath separation during use.
Source facts
- Action
- The firm, Arstasis, initiated a market withdrawal and sent notification via letters dated October 19, 2010, to customers informing the customers that Arstasis is implementing a product improvement and will replace all existing inventory that the customer has on their shelf with the improved product. The customers were instructed to work closely with their local Arstasis territory manager to ensure that their product replacement is handled satisfactorily, and complete and return the PRODUCT REPLACEMENT and PRODUCT RETURN forms to Arstasis Customer Service via fax at 1-877-688-8417 or email to customerservice@arstasis.com. If you have any questions or concerns, contact Customer Service toll free at 1-877-594-4545.
- Initiated
- 2010-10-19
- Root cause
- Device Design
- Terminated
- 2011-02-16
- Distribution
- Worldwide Distribution
- Recall number
- Z-0993-2011
- Recall status
- Terminated
- CDRH record ID
- 96420
- FDA classified
- 2011-02-04
- Recall event number
- 57507