Open original FDA record ↗Recall record
- FDA identifier
- Z-0993-2015
- Company in FDA record
- Maquet Medical Systems USA
- FDA classification date
- Jan 16, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The wrong movement results on the MAGNUS OR table. The button commands on the TEGRIS touchscreen for Lower Leg Up and Upper Leg Down are switched in the software for the integration with the MAGNUS operating table system.
Source facts
- Action
- Maquet sent an Urgent Medical Device Field Correction Letter dated August 21, 2014, to all affected customers via Federal Express. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to take the following measures until the issue is corrected: The TEGRIS equipped with a software module version up to 2.1.X.X should not be used in conjunction wiht a MAGNUS operating table system. All other MAQUET operating tables are not affected. A MAQUET service representative would contact them and arrange to have their TEGRIS software module updated. Customers with questions were instructed to contact MAQUET Technical Support at 1-888-627-8383 (press option 3, followed by option 1, then option 1 again. For questions regarding this recall call 973-709-7000.
- Initiated
- 2014-08-21
- Root cause
- Software design (manufacturing process)
- Terminated
- 2015-11-17
- Distribution
- US Nationwide Distribution including NE, NY and PA.
- Recall number
- Z-0993-2015
- Recall status
- Terminated
- CDRH record ID
- 130952
- FDA classified
- 2015-01-16
- Recall event number
- 69560