Skip to content

Public FDA record

Z-0993-2024: FilmArray Pneumonia Panel Plus (Pneumoplus), REF: RFIT-ASY-0143, CED 2797, IVD, Rx Only (30 test kit)

Recall record Z-0993-2024 for BioFire Diagnostics, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0993-2024
Company in FDA record
BioFire Diagnostics, LLC
FDA classification date
Feb 6, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Their is a potential for false positive results when using Pneumonia panel.

Source facts

Action
On December 14, 2023, Biomerieux, issued a "Urgent Medical Recall" to affected consignees via: FedEx . Biomerieux asked consignees to take the following actions: 1. Distribute this information to all appropriate personnel in your laboratory, retain a copy in your files, and forward this information to all parties that may use this product, including others to whom you may have transferred our product. 2. Use another method to confirm positive results for seasonal Coronavirus from the BIOFIRE Pneumonia Panel or BIOFIRE Pneumonia Panel plus prior to reporting the result (e.g., consider results from testing a nasopharyngeal swab with a different BIOFIRE respiratory panel such as the BIOFIRE Respiratory Panel 2.1). Alternatively, the BIOFIRE Pneumonia Panel or BIOFIRE Pneumonia Panel plus seasonal Coronavirus positive result should not be reported. 3. Complete the Acknowledgement Form in Attachment A (on the following page) and return it to bioMrieux to confirm receipt of this notice. It is important that you return the acknowledgement form to bioMrieux even if you determine that this urgent product correction notice does not impact your facility.
Initiated
2023-12-14
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Argentina, Bolivia, Brazil, Chili, China, Colombia, Costa Rica, Dominican Republic, Ecuador, Guatemala, Hong Kong, Honduras, Japan, South Korea, Mexico, Nicaragua, Panama, Peru, Philippines, Paraguay, Singapore, El Salvador, Thailand, Taiwan, Uruguay, Venezuela.
Recall number
Z-0993-2024
Recall status
Open, Classified
CDRH record ID
204923
FDA classified
2024-02-06
Recall event number
93592

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗