Lost and incorrect images. When using the "F-Rec" fluoroscopy image acquisition recording option in the "manual mode" an infrequent software error may cause images from another study to be displayed on the monitor. The acquired images may be lost requiring additional image acquisitions to be performed.
Source facts
Action
In November, 2012, Toshiba America Medical Systems, Inc. sent an Urgent: Medical Device Correction letter to customers. The letter contains a return reply form that is to be faxed to TAMS for retention.
Initiated
2012-11-08
Root cause
Software design
Terminated
2013-03-26
Distribution
USA nationwide. No international.
Recall number
Z-0994-2013
Recall status
Terminated
CDRH record ID
116745
FDA classified
2013-03-25
Recall event number
64674