Open original FDA record ↗Recall record
- FDA identifier
- Z-0994-2015
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Jan 16, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for composed images to be flipped before being sent to PACS on systems with software version VE10 and SmartOrtho license. Flipped images may be reversed so the associated annotations, e.g. labels (R/L), may be displayed incorrectly, potentially leading to misdiagnosis.
Source facts
- Action
- The firm, Siemens, sent a "Safety Advisory Notice - XP080/14/S" letter dated December 8, 2014, to end users that identified the product, problem, and actions to be taken. Siemens provided instructions to avoid the issue and indicated a software fix would be released as a permanent resolution. The customers were instructed observe this safety notice and comply with the corresponding measure until the update has been fully completed. If you have any questions, contact the Regulatory Technical Specialist at 610-219-2119.
- Initiated
- 2014-12-15
- Root cause
- Software design
- Terminated
- 2017-01-26
- Distribution
- US Distribution to states of: ID, IL, MI, MN, MO, ND and PA.
- Recall number
- Z-0994-2015
- Recall status
- Terminated
- CDRH record ID
- 132119
- FDA classified
- 2015-01-16
- Recall event number
- 70020