Z-0994-2020: Genius 2 Tympanic Thermometer Private Label Item Code: 3069

Recall record Z-0994-2020 for Cardinal Health 200, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0994-2020
Company in FDA record
Cardinal Health 200, LLC
FDA record creation date
Feb 4, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range ( 0.2C for Genius 2 and 0.3C for Genius 3 thermometers). The measurement readings drift upwards over time, which means that the thermometers could exceed the upper stated accuracy tolerance of +0.2C for Genius 2 or +0.3C for Genius 3.

Source facts

Action
Cardinal Health issued Correction Notice and Acknowledgement Form via overnight on October 28, 2019 to the direct consignees. The Medical Device Correction Notice provides instructions to our direct consignees to notify any customers who they may have further distributed affected product as to the market action. Letter states reason for recall, health risk and action to take: Customers are instructed that if they have access to a Genius Checker/Calibrator to calibrate all affected Genius thermometers. Following calibration, customers are to contact Cardinal Health Service & Repair to schedule and arrange for the software update on their Genius Checker/Calibrator. Once the updated Genius Checker/Calibrator has been returned to your facility, recalibrate all thermometers. If a customer does not have access to a Genius Checker/Calibrator, they are instructed to contact Cardinal Health Service & Repair to schedule and arrange for their thermometer(s) to be sent to a service center. Contact Cardinal Health Service & Repair to schedule and arrange for your thermometer(s) to be sent to one of our service centers. Monday Friday between 8:00am - 8:00pm EST Service and Repair Line 877-227-3462, Option 1
Initiated
2019-10-28
Root cause
Under Investigation by firm
Terminated
2022-05-17
Distribution
Worldwide Distribution - US Nationwide and Foreign in the countries of: Australia Austria Belgium Canada Canary Islands Chile Costa Rica Croatia Cyprus Czech Republic Denmark Egypt Estonia Finland France French Guiana French Polynesia Gabon Germany Gibraltar Greece Guadeloupe Hong Kong Iran, Islamic Republic Of Ireland Israel Italy Kuwait Liechtenstein Luxembourg Martinique Mayotte Mexico Netherlands New Caledonia New Zealand Norway Panama Philippines Poland Portugal Puerto Rico Qatar Reunion Romania San Marino Saudi Arabia Singapore Slovakia Slovenia South Africa South Korea Spain Swaziland Sweden Switzerland Taiwan Thailand Turkey United Arab Emirates United Kingdom
Recall number
Z-0994-2020
Recall status
Terminated
CDRH record ID
177651
FDA record created
2020-02-04
Recall event number
84202

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗