Z-0995-2020: GORE EXCLUDER AAA Endoprosthesis

Recall record Z-0995-2020 for W. L. Gore & Associates Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0995-2020
Company in FDA record
W. L. Gore & Associates Inc.
FDA record creation date
Feb 4, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Firm has received reports of leading end catheter component separations.

Source facts

Action
The recall consists of corrections with no proposed product removal. The correction will consist of: A Medical Device Safety Correction letter sent to US physicians and hospitals and OUS hospitals and Distributors, along with an IFU Summary of Changes document Updates to the products Instruction for Use Updates to physician didactic training Additional manufacturing process error-proofing The firm is also instructing customers to share the letter with others in their hospital or clinic as appropriate, and contact the firm's Customer Service (email: MPDCustomerCare@wlgore.com or by phone at 800-528- 8763) with any questions related to the letter. Any adverse event should be reported to the manufacturer and the country specific regulatory authorities immediately. To report an event to Recalling Firm, email: medcomplaints@wlgore.com or contact: USA: +1.800.528.1866, Ext. 44922, +1.928.864.4922, Fax: +1.928.864.4364 China: +86 21 5172 8237, Fax: +86 21 5172 8236 Japan: +81 3 6746 2562, Fax: +81 3 6746 2563 Brazil: +55 11 5502-7953, Fax: +55 11 5502-7965 EMEA: +49 89 4612 3440, Fax: +49 89 4612 43440
Initiated
2020-01-06
Root cause
Employee error
Terminated
2022-08-09
Distribution
US: OUS: Australia; Austria; Canada; China; Taiwan; United Arab Emirates; Argentina; Brazil; Switzerland; Chile; Colombia; Costa Rica; Czechia; Spain; Greece; Croatia; Hungary; Ireland; Israel; Italy; United Kingdom; Lebanon; Mexico; New Zealand; Panama; Poland; Saudi Arabia; Slovenia; Germany; Netherlands; Norway; Finland; France; Sweden; Monaco;
Recall number
Z-0995-2020
Recall status
Terminated
CDRH record ID
178536
FDA record created
2020-02-04
Recall event number
84533

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗