Open original FDA record ↗Recall record
- FDA identifier
- Z-0995-2020
- Company in FDA record
- W. L. Gore & Associates Inc.
- FDA record creation date
- Feb 4, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Firm has received reports of leading end catheter component separations.
Source facts
- Action
- The recall consists of corrections with no proposed product removal. The correction will consist of: A Medical Device Safety Correction letter sent to US physicians and hospitals and OUS hospitals and Distributors, along with an IFU Summary of Changes document Updates to the products Instruction for Use Updates to physician didactic training Additional manufacturing process error-proofing The firm is also instructing customers to share the letter with others in their hospital or clinic as appropriate, and contact the firm's Customer Service (email: MPDCustomerCare@wlgore.com or by phone at 800-528- 8763) with any questions related to the letter. Any adverse event should be reported to the manufacturer and the country specific regulatory authorities immediately. To report an event to Recalling Firm, email: medcomplaints@wlgore.com or contact: USA: +1.800.528.1866, Ext. 44922, +1.928.864.4922, Fax: +1.928.864.4364 China: +86 21 5172 8237, Fax: +86 21 5172 8236 Japan: +81 3 6746 2562, Fax: +81 3 6746 2563 Brazil: +55 11 5502-7953, Fax: +55 11 5502-7965 EMEA: +49 89 4612 3440, Fax: +49 89 4612 43440
- Initiated
- 2020-01-06
- Root cause
- Employee error
- Terminated
- 2022-08-09
- Distribution
- US: OUS: Australia; Austria; Canada; China; Taiwan; United Arab Emirates; Argentina; Brazil; Switzerland; Chile; Colombia; Costa Rica; Czechia; Spain; Greece; Croatia; Hungary; Ireland; Israel; Italy; United Kingdom; Lebanon; Mexico; New Zealand; Panama; Poland; Saudi Arabia; Slovenia; Germany; Netherlands; Norway; Finland; France; Sweden; Monaco;
- Recall number
- Z-0995-2020
- Recall status
- Terminated
- CDRH record ID
- 178536
- FDA record created
- 2020-02-04
- Recall event number
- 84533